Product Catalog: |
Magnetic Field Coil
Product Detail: |
DXMC-1MF Helmholtz Coil
Product Introduction:
Typical uniaxial Helmholtz coil is composed two coils with the same number of turns,same winding method, radius equal to the coil spacing. DXMC- MF series Helmholtz coils usually is used to produce magnetic field whose designated volum is relatively large, uniformity is relatively high, but the value of the magnetic field is weak. Users can use this field to complete all kinds of experiments and the simulation of weak magnetic field. Its main application:to offset the earth magnetic field,to judge magnetic shielding effect, magnetic coefficient of electronic equipments, calibration for fluxgate magneticmeter and navigational equipments, study of biological magnetic field.
Specification for example: DXMC-1MF36 Helmholtz coil,magnetic field 0-18.5Gs.
Routine constants of coil: K=0.2023 (cm), Physical radius of coill R=181mm.
wire Diameter |
inner diameter of coil |
external diameter of coil |
turns off unicoil |
coil resistance (20℃) |
1.0mm |
340mm |
384mm |
125 |
6.350Ω |
Test for magnetic field(spacing between the centres of two coils),formula of magnetic field coresponding current: H=6.2033I(Gs)
Constant current (A) |
Magntic field calculating data(Oe) |
Actual test data for magnetic field distribution in uniform area (Gs) |
||||
(-30,-30) |
(-30,30) |
(30,30) |
(30,-30) |
(0,0) |
||
3.000 |
18.609 |
18.64 |
18.64 |
18.63 |
18.64 |
18.61 |
2.000 |
12.407 |
12.43 |
12.43 |
12.42 |
12.41 |
12.41 |
1.500 |
9.305 |
9.32 |
9.32 |
9.31 |
9.31 |
9.30 |
1.000 |
6.203 |
6.22 |
6.22 |
6.21 |
6.21 |
6.20 |
0.750 |
4.652 |
4.67 |
4.67 |
4.66 |
4.66 |
4.7 |
0.500 |
3.102 |
3.11 |
3.11 |
3.11 |
3.11 |
3.10 |
0.250 |
1.551 |
1.56 |
1.56 |
1.55 |
1.55 |
1.55 |
Linearity of actual coil: 0.999%, actual use could set according to the formula of magnetic field coresponding current. |
NOTE:
1. the coil allowed max.working current:1.000A(without cooling within 24h),linearity of coil:superior to 0.5%
2. the coil allowed max.working current:2.000A(without cooling within 1h)
3.the coil allowed max.working current:3.000A(without cooling within 10min)
The company designs and produces Helmholtz coil professionally and quickly, the processing cycle for one-dimensional magnetic field coil or one-dimensional test coil is 10 days.