Product Detail: |
DX-330F High Resolution Three Axis Fluxgate Magnetometer
Overview
Dexing fluxgate magnetometers are instruments for accurate measurement of weak static and dynamic and low frequency vector magnetic fields. Compared with a magnetic field measuring instrument with high stability, linearity and accuracy, it is compared with the principle of Hall effect or the principle of magneto-resistance effect. It is the best choice to measure the weak magnetic field, with the output of digital and analog signals , the unique CAN bus network interface is suitable for multi - array magnetic field testing , and is widely used in the fields of scientific research , military industry , aerospace , etc .
Hardware interface information of Flux-gate Magnetometer
Interface DB15 |
Probe input/output |
Measuring mode |
Hand-held probe or homemade tooling |
Reading mode |
Display screen or computer software |
Display |
7-bit SVG color screen |
BNC |
Analog signal output |
Power supply |
220V |
RS-232 |
Digital signal input / output (providing instruction format) |
Parameter
DX-330F |
||||
Size(mm)(width * height * depth) |
320*130*285 |
|||
Dimensions |
3 |
|||
Resolution |
0.1nT |
|||
Preheating Time |
15mins |
|||
Accuracy |
< 0.5% of reading |
|||
Noise level |
10 -20pTrms/√Hz at 1Hz |
|||
Range (optional) |
±100μT |
±500μT |
±1000μT |
|
Error of temperature coefficient |
1nT/ºC |
±10nT/°C |
±10nT/°C |
|
Bandwidth at -3db |
>1.5kHz |
>100Hz |
||
Three-dimensional orthogonality error |
<1° |
<2° |
||
Frequency response (DC deviation: < 5) |
1000Hz |
1000Hz |
||
Operating temperature range |
-40ºC to +70ºC |
0°C to +215°C |
||
Acquisition rate |
0-300 times/second |
0-300 times/second |
||
Probe size(mm) (original British) |
32*32*122 |
23*23*21.4 |
||
Communication |
RS-232/USB |
RS-232/USB |
||
Analog output |
3 channels BNC |
3 channels BNC |
||
Supporting software |
Data reading software |
Data reading software |
Backpanel
- Power Interface and Switch
- Rs-232c Connector
- X Channel Analog Output BNC
- Y Channel Analog Output BNC
- Z Channel Analog Output BNC
- Probe Input Connector
Applied range:
- Monitoring the magnetic field of earth environment, laboratory, and space.
- Calibrating laboratory magnetic field shielding and generating field sources, such as: 3D Helmholtz coil system, non-distance coil system etc.
- Geological prospecting. Petroleum system while drilling system detect weak magnetic field in rock.
- Testing the attenuation characteristics and shielding effectiveness of magnetic field shielding system.
- Detecting magnetic shielding effect in magnetic shielding room.
- Dipole moment and remanence magnetic field of satellite.
- Calibration standard for flux-gate meter, magneto-optical effect magnetometer and all kinds of sensors.
- Earthquake precursor monitoring, volcanic observation, and other environmental and disaster geological work.
- Aeronautical and marine magnetic measurements.
- Research on weak magnetic field in bioscience.