Teslameter,Gaussmeter,Magnetometers,Portable gaussmeter,Gauss meter,Tesla meter logo
Teslameter,Gaussmeter,Magnetometers,Portable gaussmeter,Gauss meter,Tesla meter
Product Catalog:

Digital Two(Three) Dimensional Gauss / Tesla Meter

Product Overview:

DX-210 Digital Two Dimensional Gauss / Tesla Meter

DX-210 Gauss / Tesla Meter Detail:

DC Measurement(DC):

DC display resolution 5 ¾ (6 digital filter display)  

DX-1800 *0-3mT 0-30mT 0-300mT 0-3T
Resolution 0.00001mT 0.0001mT 0.001mT 0.01mT

Marked * is the Automatic identification rangs of low magnetic field probe
DC Accuracy: ±0.15% of reading ±0.05% of range
DC Temperature Coefficient:±(0.03% ±1 count)/ ºC
Acquisition Speed: 20/s

AC Effective Values(AC) and Frequency Measurement
AC Display Resolution: 5 ¾ (6 digital filter display)
AC RMS Frequency Range: 20Hz-2500Hz
AC RMS Accuracy: ±2% of reading
AC RMS Frequency Response: 20Hz-2500Hz
Acquisition Speed: 20/s

Front Panel

Display: 800 x480 Pixel,7 inch full view color screen
Display Resolution: 5 ¾ (6 digital filter display)
Display Update Rate: 5 rdg /s
Display Units: Gauss(G), Tesla(T), Amps/m(A/m), Oster(Oe)
Unit Multiplier: m,k,M
Definitions of control keys: AC/DC Switch, Measurement Units, Range Switch, Data Storage, Up/Down/Left/Right Direction, Menu, Max/Min/Interface Lock, Zero Mode, ConfirmationButton, Relative Mode, Reset Mode
Keyboard: 12 membrane keys
Front Panel Features: Good Communication Ability, Intuitive Operation, Display Hints, Brightness Control, Panel Lock

Probe Specifications

Input Type: Single Hall Effect Two Dimensional Sensor
Probe Features: High linear digital Hall probe, Probe Auto Zero
Probe Compatibility: Standard probe or special probe  
Connector: 15 pin D-type

Interface Features

RS - 232C interface function  
Baudrate: 19600, 57600, 115200
Connector: 9 pin D-type, Terminal Configuration    
Software Suppot: Standard interface definition
Analog output monitoring
Configuration: Real-time analog voltage output
Rang: ± 3 V voltage
Frequency Response: DC to 2500Hz
Accuracy: depending on probes
Minimum load resistance: 1 k (Short-Circuit Protection)
Connector: BNC

General Parameters

Chassis: Metal shielded
Environment Temperature: 15℃to 35℃rated accuracy; 5℃to 40℃accuracy reducing                                                                                                      
Power Requirements: 220-240(+5 %, -10 %), 50 or 60Hz
Size: 320 mmWidth× 110mmheight× 287mm
Weight:4kg(7.8lbs)

 

Optional probes

*The length could be customized as requirements           
*We can provide the probes with temperature sesors       
* May get gilded outside     
* Lead time is about 1-4 week          

Model
L(mm)
apporx
D(mm)
A(mm)
Active area
(mm2)
Frequency range
Full Scale
magnetic linearity(%rdg)
operating temp range
Zero field offset voltage(mv)
Temp
Coefficient
(max)
stem material
2DXP801F
10/25/50
4.5
1±0.1
0.15
DC-800HZ
30KG
±0.2%to 30kG
-10°Ct0+75°C
0.01to0.2
±0.03%/C°
Brass
2DXP802F
10/25/50
3.5
1±0.1
015
DC-800HZ
30KG
±0.2%to 30kG
-10°Ct0+75°C
0.01to0.2
±0.03%/C°
Brass
2DXP803F
10/25/50
2.0
1±0.1
0.15
DC-800HZ
30KG
±0.2%to 30kG
-10°Ct0+75°C
0.01to0.2
±0.03%/C°°
Brass
2DXP804F
10/25/50
1.5
1±0.1
0.15
DC-800HZ
30KG
±0.2%to 30kG
-10°Ct0+75°C
0.01to0.2
±0.03%/C°
Brass
2DXP805F
10/25/50
1.2
1±0.1
0.10
DC-800HZ
30KG
±0.2%to 30kG
-10°Ct0+75°C
0.01to0.2
±0.03%/C°
Brass
2DXP806F
10/25/50
0.9
1±0.1
0.10
DC-800HZ
30KG
±0.2%to 30kG
-10°Ct0+75°C
0.01to0.2
±0.03%/C°
Brass

Transverse probes

*The length could be customized as requirements           
*We can provide the probes with temperature sesors       
* May get gilded outside     
* Lead time is about 1-4 week      

Model
L(cm)
approx
T
(mm)
W
(mm)
A
(mm)
Active area
(mm2)
Frequency range
Full Scale (Standard)
magnetic linearity
(%rdg)
operating temp range
Zero field offset voltage(mv)
TemP
Coefficient
(max)
stem material
DCG801F
4.5/7
0.78
2
1±0.1
0.15
DC-400Hz
30KG
±0.2% to30kG
-10°Ct0+75°C
0.01-0.2
±0.03%/C°
Brass
DXG802F
4.5/7
0.62
1.5
1±0.1
0.15
DC-400HZ
30KG
±0.2% to30kG
-10°Ct0+75°C
0.01-0.2
±0.03%/C°
Brass
DXD801F
4.5/7
0.50
1.68
1±0.1
0.15
DC-400HZ
30KG
±0.2% to30kG
-10°Ct0+75°C
0.01-0.2
±0.03%/C°
Brass
DXD802F
4.5/7
0.35
0.7
0.5±0.1
0.15
DC-400HZ
30KG
±0.2% to30kG
-10°Ct0+75°C
0.01-0.2
±0.03%/C°
Resin
DXD803F
4.5/7
0.2
07
0.5±0.1
0.10
DC-400HZ
30KG
±0.2% to30kG
-10°Ct0+75°C
0.01-0.2
±0.03%/C°
Resin

 

Axial probes

*The length could be customized as requirements           
*We can provide the probes with temperature sesors       
* May get gilded outside     
* Lead time is about 1-4 week             

Model
L(mm)
apporx
D(mm)
A(mm)
Active area
(mm2)
Frequency range
Full Scale
magnetic linearity(%rdg)
operating temp range
Zero field offset voltage(mv)
Temp
Coefficient
(max)
stem material
1DXP801F
10/25/50
3.5
1±0.1
015
DC-800HZ
30KG
±0.2%to 30kG
-10°Ct0+75°C
0.01to0.2
±0.03%/C°
Brass
1DXP803F
10/25/50
2.0
1±0.1
0.15
DC-800HZ
30KG
±0.2%to 30kG
-10°Ct0+75°C
0.01to0.2
±0.03%/C°°
Brass
1DXP803F
10/25/50
1.5
1±0.1
0.15
DC-800HZ
30KG
±0.2%to 30kG
-10°Ct0+75°C
0.01to0.2
±0.03%/C°
Brass
1DXP804F
10/25/50
1.2
1±0.1
0.10
DC-800HZ
30KG
±0.2%to 30kG
-10°Ct0+75°C
0.01to0.2
±0.03%/C°
Brass
1DXP805F
10/25/50
0.9
1±0.1
0.10
DC-800HZ
30KG
±0.2%to 30kG
-10°Ct0+75°C
0.01to0.2
±0.03%/C°
Brass