Product Catalog: |
Teslameter / Gaussmeter
- DX Teslameter / Gaussmeter
- DX-150 Intelligent Digital Gauss Meter / Tesla Meter
- DX-160 Intelligent Digital Gauss Meter / Tesla Meter
- DX-180 Intelligent Digital Gauss Meter / Tesla Meter
Product Details: |
DX-180 Intelligent Digital Gauss Meter / Tesla Meter
DX-180 Gauss/Telsa Meter Details:
DC Measurement(DC):
DC display resolution: 5 ¾ (6 digital filter display)
DX-180 Range |
*0-3mT |
0-30mT |
0-300mT |
0-3T |
*0-10T |
Resolution |
0.00001mT |
0.0001mT |
0.001mT |
0.01mT |
0.1mT |
Marked * are the automatic identification rangs of low magnetic field probe and ultra-high magnetic field probe
DC Accuracy: ±0.05% of reading ±0.005% of range
DC Temperature Coefficient: ±(0.01% ±1 count)/ ºC
Acquisition Speed: 20 times/sec
AC Effective Values(AC) and Frequency Measurement
AC Resolution: 0.01mT
AC RMS Frequency Range: 2Hz-50kHz
AC RMS Accuracy: ±1% of reading ±0.3% of range
AC RMS Frequency Response: 20Hz-2500Hz
Acquisition Speed: 15 times/sec
Front Panel
Display: 320 x 240 pixel, 3.5 inch full view color screen
Display Resolution: 5 ¾ (6 bit digital filter display)
Display Update Rate: 5 rdg /sec
Display Units: Gauss(G), Tesla(T), Amps/m(A/m), Oster(Oe)
Unit Multiplier: m, k, M
Definitions of control keys: AC/DC Switch, Measurement Units, Range Switch, Data Storage, Up/Down/Left/Right Direction, Menu, Max/Min/Interface Lock, Zero Mode, Confirmation Button, Relative Mode, Reset Mode
Keyboard: 14 membrane keys
Front Panel Features: Good Communication Ability, Intuitive Operation, Display Hints, Brightness Control, Panel Lock, Graphic Display
Probe Specifications
Input Type: Single Hall Effect Sensor
Probe Features: High linear digital Hall probe, Probe Auto Zero, Temperature Compensation, Ultra-thin only 0.80mm
Probe Compatibility: Standard probe or special probe
Connector: 15 pin D-type
Interface Features
RS - 232C interface function
Baudrate: 19600, 57600, 115200
Connector: 9 pin D-type, Terminal configuration
Software Support: Standard interface definition
Analog output monitoring
Configuration: Real-time analog voltage output
Rang: ± 3 V
Frequency Response: DC to 2500Hz
Accuracy: depending on probes
Minimum load resistance: 1 k (Short-Circuit Protection)
Connector: BNC
General Parameters
Chassis: Metal shielded
Environment Temperature: 15℃to 35℃rated accuracy; 5℃to 40℃accuracy reducing
Power Requirements: 220-240V (+5%, -10%), 50 or 60Hz
Size: L287mm × H110mm × W320 mm
Weight: 4kg(7.8lbs)
Optional probe:
Transverse probes
*The length and shape of probes could be customized as requirements
*We can provide the probes with temperature sensors
* May get gilded outside
* Lead time is about 1-4 week
Model |
L(cm) approx |
T (mm) max |
W (mm) |
A (mm) |
Active area (mm) |
Frequency range |
Full Scale (Standard) |
Magnetic linearity(%rdg) |
Operating temp range |
Zero temp coefficient (max) |
Temp coefficient (max) |
Stem material |
HCDX800F |
6/8/10 |
1.00 |
2.6 |
1±0.1 |
0.15 |
DC-30kHZ |
100kG |
±0.05%to30kG |
-10°Cto+100°C |
±0.08 G/ºC |
±0.03%/C° |
Brass |
HCDX801F |
6/8/10 |
0.80 |
2.2 |
1±0.1 |
0.15 |
DC-30kHZ |
100kG |
±0.05%to30kG |
-10°Cto+100°C |
±0.08 G/ºC |
±0.03%/C° |
Brass |
HCDX802F |
6/8/10 |
0.60 |
1.55 |
1±0.1 |
0.15 |
DC-30kHZ |
100kG |
±0.05%to30kG |
-10°Cto+100°C |
±0.08 G/ºC |
±0.03%/C° |
Brass |
HMDX801F |
6/8/10 |
0.50 |
1.7 |
1±0.1 |
0.15 |
DC-30kHZ |
100kG |
±0.05%to30kG |
-10°Cto+100°C |
±0.08 G/ºC |
±0.03%/C° |
Brass |
HMDX802F Flexible |
6/8/10 |
0.35 |
1.2 |
0.5±0.1 |
0.15 |
DC-30kHZ |
100kG |
±0.2%to30kG |
-10°Cto+100°C |
±0.10 G/ºC |
±0.03%/C° |
Resin |
HMDX803F Flexible |
6/8/10 |
0.25 |
0.7 |
0.5±0.1 |
0.10 |
DC-30kHZ |
100kG |
±0.5%to30kG |
-10°Cto+100°C |
±0.10 G/ºC |
±0.03%/C° |
Resin |
The model in red color is standard probe.
Axial probes
*The length and shape of probes could be customized as requirements
*We can provide the probes with temperature sensors
* May get gilded outside
* Lead time is about 1-4 week
Model |
L(cm) approx |
D(mm) max |
A (mm) |
Active area (mm) |
Frequency range |
Full Scale (Standard) |
Magnetic linearity (%rdg) |
Operating temp range |
Zero temp coefficient (max) |
Temp coefficient (max) |
Stem material |
1HADX800F |
10/25/50 |
5.00 |
1±0.1 |
0.15 |
DC-30KHZ |
300kG |
±0.05%to30kG |
-10°Cto+100°C |
±0.08 G/ºC |
±0.03%/C° |
Brass |
1HADX801F |
10/25/50 |
3.00 |
1±0.1 |
0.15 |
DC-30KHZ |
300kG |
±0.05%to30kG |
-10°Cto+100°C |
±0.08 G/ºC |
±0.03%/C° |
Brass |
1HADX802F |
10/25/50 |
2.00 |
1±0.1 |
0.15 |
DC-30KHZ |
300kG |
±0.05%to30kG |
-10°Cto+100°C |
±0.08 G/ºC |
±0.03%/C° |
Brass |
1HADX803F |
10/25/50 |
1.50 |
1±0.1 |
0.15 |
DC-30KHZ |
300kG |
±0.05%to30kG |
-10°Cto+100°C |
±0.08 G/ºC |
±0.03%/C° |
Brass |
1HADX804F |
10/25/50 |
1.20 |
0.5±0.1 |
0.15 |
DC-30KHZ |
300kG |
±0.05%to30kG |
-10°Cto+100°C |
±0.08 G/ºC |
±0.03%/C° |
Brass |
1HADX805F 10/25/50 |
10/25/50 |
0.90 |
0.5±0.1 |
0.10 |
DC-30KHZ |
300kG |
±0.5%to30kG |
-10°Cto+100°C |
±0.12 G/ºC |
±0.03%/C° |
Brass |